Characterization and potential for reducing optical resonances in Fourier transform infrared spectrometers of the Network for the Detection of Atmospheric Composition Change (NDACC)
Although optical components in Fourier transform infrared (FTIR) spectrometers are preferably wedged, in practice, infrared spectra typically suffer from the effects of optical resonances ("channeling") affecting the retrieval of weakly absorbing gases. This study investigates the level of channeling of each FTIR spectrometer within the Network for the Detection of Atmospheric Composition Change (NDACC). Dedicated spectra were recorded by more than 20 NDACC FTIR spectrometers using a laboratory mid-infrared source and two detectors. In the indium antimonide (InSb) detector domain (1900â5000âcmâ1), we found that the amplitude of the most pronounced channeling frequency amounts to 0.1ââ° to 2.0ââ° of the spectral background level, with a mean of (0.68±0.48)ââ° and a median of 0.60ââ°. In the mercury cadmium telluride (HgCdTe) detector domain (700â1300âcmâ1), we find even stronger effects, with the largest amplitude ranging from 0.3ââ° to 21ââ° with a mean of (2.45±4.50)ââ° and a median of 1.2ââ°. For both detectors, the leading channeling frequencies are 0.9 and 0.11 or 0.23âcmâ1 in most spectrometers. The observed spectral frequencies of 0.11 and 0.23âcmâ1 correspond to the optical thickness of the beam splitter substrate. The 0.9âcmâ1 channeling is caused by the air gap in between the beam splitter and compensator plate. Since the air gap is a significant source of channeling and the corresponding amplitude differs strongly between spectrometers, we propose new beam splitters with the wedge of the air gap increased to at least 0.8â. We tested the insertion of spacers in a beam splitter's air gap to demonstrate that increasing the wedge of the air gap decreases the 0.9âcmâ1 channeling amplitude significantly. A wedge of the air gap of 0.8â reduces the channeling amplitude by about 50â%, while a wedge of about 2â removes the 0.9âcmâ1 channeling completely. This study shows the potential for reducing channeling in the FTIR spectrometers operated by the NDACC, thereby increasing the quality of recorded spectra across the network.
document
https://n2t.org/ark:/85065/d7h135bt
eng
geoscientificInformation
Text
publication
2016-01-01T00:00:00Z
publication
2021-02-17T00:00:00Z
Copyright author(s). This work is licensed under a Creative Commons Attribution-NonCommercial 4.0 International License.
None
OpenSky Support
UCAR/NCAR - Library
PO Box 3000
Boulder
80307-3000
name: homepage
pointOfContact
OpenSky Support
UCAR/NCAR - Library
PO Box 3000
Boulder
80307-3000
name: homepage
pointOfContact
2025-12-25T06:34:44.886896